Autosampler vial sizes and compatibility
Updated
Most HPLC autosampler problems with vials come from mixing formats: a cap made for one thread on a vial made for another, an insert too long for the needle depth, or a septum of the wrong diameter. This guide sets out the standard formats using manufacturer dimensions. For choosing septa material and glass type see the vials, caps & septa guide.
Standard 2 mL vial formats
The common autosampler vial is 2 mL, 12 x 32 mm[3]. It comes with three closure styles, and each needs its own septum size[1, 2]:
| Closure | Septum / seat diameter | Cap outer diameter | Cap opening |
|---|---|---|---|
| Screw, wide opening (9 mm, 9-425) | 9 mm | 12 mm | 6 mm |
| Screw, short opening (8-425) | 8 mm | 12 mm | 5 mm |
| Crimp, wide opening | 11 mm | 11 mm | 6 mm |
| Snap, wide opening | 11 mm | 12 mm | 6 mm |
Source: Agilent[1]. A 9 mm cap will not seal an 8-425 vial and vice versa.
Vials and caps in the catalog by format
- 9 mm screw, 2 mL: Agilent 2 mL Clear Screw-Top Vials, Certified, 100/pk (
5182-0714) (Agilent lists it for Agilent, Waters, Thermo, Shimadzu, PerkinElmer and Varian/Bruker instruments[3]), Agilent 2 mL Clear Screw-Top Vials with Write-On Spot, Certified, 100/pk (5182-0715), Agilent 2 mL Amber Screw-Top Vials with Write-On Spot, Certified, 100/pk (5182-0716); caps Agilent Blue Screw Caps with PTFE/Red Silicone Septa, 100/pk (5182-0717) and Agilent Blue Screw Caps, Bonded Pre-Slit PTFE/Silicone Septa, Certified, 100/pk (5185-5824); replacement septa Agilent 9 mm PTFE/Red Silicone Septa for Screw Caps, Certified, 100/pk (5182-0731). - 12 x 32 mm screw-neck kits: Waters LCGC Certified 12 x 32 mm Clear Screw-Neck Vials with Pre-Slit Caps, 100/pk (
186000307C), Waters LCGC Certified Total Recovery Vials, 12 x 32 mm, 100/pk (186000384C), Thermo Scientific SureSTART 2 mL Clear Screw Vial Kit, Level 3, Bonded Slit Septa, 100/pk (6PMSCK641W). - 8-425 caps: Thermo Scientific White Screw Caps with Septa for Shimadzu Autosamplers, 8-425, 100/pk (
C4013-63W). - 1.5 mL vials (Shimadzu and short-thread): Shimadzu Certified Vial Kit for LC/LC-MS, 1.5 mL (100/pk) (
227-34001-01), Shimadzu Screw Vial Kit, Clear Glass, 1.5 mL (100/pk) (961-10010-06), Shimadzu Shim-vial H Low-Adsorption Certified Vial Kit, 1.5 mL (100/pk) (227-34500-01), Shimadzu Vial Kit for PFAS Analysis, Clear Plastic, 1.5 mL (100/pk) (961-10010-34) and the compatible Altmann Analytik 1.5 mL Short-Thread Certified Vial Kit with UltraClean Seal, 100/kit (AAV22513667) (ND9 short thread).
Fill volume and inserts
Agilent gives a recommended fill of 1.5 mL for the 2 mL screw vial 5182-0714[2]. For small samples, inserts reduce the volume needed: Agilent lists Agilent 250 µL Glass Inserts with Polymer Feet, 100/pk (5181-1270) as 30 x 5.6 mm and Agilent 250 µL Pulled-Point Glass Vial Inserts, 100/pk (5183-2085) as 31 x 5.6 mm, both 250 µL[2]. Macherey-Nagel 0.2 mL Conical Glass Insert for Wide-Opening Vials, 100/pk (702813) (6 mm OD x 31 mm, per its listing) fits wide-opening vials, and Shimadzu Micro Insert, Clear Glass, 0.1 mL for 1.5 mL Vials (100/pk) (961-10020-27) is made for 1.5 mL Shimadzu vials. Check the needle-depth setting of your autosampler when you start using inserts.
Before you switch vial suppliers
- Confirm vial body (12 x 32 mm), thread or closure style and septum diameter match your current vial.
- Check your autosampler manual for supported vial types and needle depth.
- Run a few blank injections to confirm there is no septum coring or carry-over.
Filter samples before they go into the vial: see the syringe filter selection guide. Browse all vials, caps & inserts.
Frequently asked questions
What is the difference between 9 mm and 8-425 screw vials?
Both are 2 mL, 12 x 32 mm vials. Agilent specifies the wide-opening 9 mm screw closure with a 9 mm septum and a 6 mm cap opening, and the short-opening 8-425 screw closure with an 8 mm septum and a 5 mm cap opening. Caps and septa are not interchangeable between the two.
Will Agilent 2 mL vials fit other brands of autosampler?
Agilent lists its 5182-0714 2 mL screw vial as compatible with Agilent, Waters, Thermo, Shimadzu, PerkinElmer and Varian/Bruker instruments. Always confirm against your own autosampler manual, especially needle depth when using inserts.
How full should I fill a 2 mL vial?
Agilent gives a recommended fill of 1.5 mL for its 2 mL screw vial 5182-0714, and 250 µL for its 250 µL inserts. Overfilling can cause septum contact and carry-over.