Autosampler vial sizes and compatibility

Updated

Most HPLC autosampler problems with vials come from mixing formats: a cap made for one thread on a vial made for another, an insert too long for the needle depth, or a septum of the wrong diameter. This guide sets out the standard formats using manufacturer dimensions. For choosing septa material and glass type see the vials, caps & septa guide.

Standard 2 mL vial formats

The common autosampler vial is 2 mL, 12 x 32 mm[3]. It comes with three closure styles, and each needs its own septum size[1, 2]:

2 mL vial closures (Agilent dimensions)
ClosureSeptum / seat diameterCap outer diameterCap opening
Screw, wide opening (9 mm, 9-425)9 mm12 mm6 mm
Screw, short opening (8-425)8 mm12 mm5 mm
Crimp, wide opening11 mm11 mm6 mm
Snap, wide opening11 mm12 mm6 mm

Source: Agilent[1]. A 9 mm cap will not seal an 8-425 vial and vice versa.

Vials and caps in the catalog by format

Fill volume and inserts

Agilent gives a recommended fill of 1.5 mL for the 2 mL screw vial 5182-0714[2]. For small samples, inserts reduce the volume needed: Agilent lists Agilent 250 µL Glass Inserts with Polymer Feet, 100/pk (5181-1270) as 30 x 5.6 mm and Agilent 250 µL Pulled-Point Glass Vial Inserts, 100/pk (5183-2085) as 31 x 5.6 mm, both 250 µL[2]. Macherey-Nagel 0.2 mL Conical Glass Insert for Wide-Opening Vials, 100/pk (702813) (6 mm OD x 31 mm, per its listing) fits wide-opening vials, and Shimadzu Micro Insert, Clear Glass, 0.1 mL for 1.5 mL Vials (100/pk) (961-10020-27) is made for 1.5 mL Shimadzu vials. Check the needle-depth setting of your autosampler when you start using inserts.

Before you switch vial suppliers

  1. Confirm vial body (12 x 32 mm), thread or closure style and septum diameter match your current vial.
  2. Check your autosampler manual for supported vial types and needle depth.
  3. Run a few blank injections to confirm there is no septum coring or carry-over.

Filter samples before they go into the vial: see the syringe filter selection guide. Browse all vials, caps & inserts.

Frequently asked questions

What is the difference between 9 mm and 8-425 screw vials?

Both are 2 mL, 12 x 32 mm vials. Agilent specifies the wide-opening 9 mm screw closure with a 9 mm septum and a 6 mm cap opening, and the short-opening 8-425 screw closure with an 8 mm septum and a 5 mm cap opening. Caps and septa are not interchangeable between the two.

Will Agilent 2 mL vials fit other brands of autosampler?

Agilent lists its 5182-0714 2 mL screw vial as compatible with Agilent, Waters, Thermo, Shimadzu, PerkinElmer and Varian/Bruker instruments. Always confirm against your own autosampler manual, especially needle depth when using inserts.

How full should I fill a 2 mL vial?

Agilent gives a recommended fill of 1.5 mL for its 2 mL screw vial 5182-0714, and 250 µL for its 250 µL inserts. Overfilling can cause septum contact and carry-over.

Sources

  1. Agilent: Autosampler Vials, Caps, and Septa Basics (vial and closure dimensions)
  2. Agilent: Vials and Sample Containment Solutions catalog (5994-4803EN)
  3. Agilent: 5182-0714 screw top vial – specifications