How to choose an HPLC column
Updated
Choosing a column is mostly a sequence of five decisions. Get the first two right and the rest is optimisation.
1. Pick the separation mode from your analyte
- Small, neutral to moderately polar molecules soluble in water/organic mixtures → reversed phase (C18, C8, phenyl, PFP). This covers most pharmaceutical, environmental and food methods.
- Very polar compounds that elute at the void on C18 (metabolites, amino acids, nucleosides) → HILIC, e.g. Waters ACQUITY UPLC BEH Amide Column, 2.1 x 100 mm, 1.7 µm (
186004801) or Shodex HILICpak VG-50 4E HILIC Column, 4.6 x 250 mm, 5 µm (F7630100). - Enantiomers → chiral columns such as Daicel CHIRALPAK AD-H Chiral HPLC Column, 4.6 x 250 mm, 5 µm (
19325) or the solvent-robust immobilised Daicel CHIRALPAK IA Immobilized Chiral Column, 4.6 x 250 mm, 5 µm (80325). - Polymers by molecular weight → SEC/GPC, e.g. Agilent PLgel MIXED-C GPC/SEC Column, 7.5 x 300 mm, 5 µm (
PL1110-6500) or Shodex GPC KF-806M Mixed-Bed GPC Column, 8.0 x 300 mm, 10 µm (F6028090); proteins → aqueous SEC such as Agilent AdvanceBio SEC 300 Å Protein Column, 7.8 x 300 mm, 2.7 µm (PL1180-5301). - Sugars and organic acids → ligand-exchange / ion-exclusion columns (see the comparison).
2. Choose the stationary phase
Start with a well-end-capped, high-purity C18 (USP L1). Move to a C8 (L7) for less retention, a phenyl or biphenyl (e.g. Restek Raptor Inert Biphenyl Column, 4.6 x 150 mm, 2.7 µm (9309A65-T)) or PFP for aromatic and positional-isomer selectivity, and a polar-embedded or aqueous-stable C18 for highly aqueous conditions. If your method must run at high pH, choose a hybrid or pH-robust particle such as Waters XBridge C18 HPLC Column, 4.6 x 150 mm, 5 µm, 130 Å (186003116) (pH 1–12 per source specification).
3. Choose dimensions
- 4.6 mm ID — standard HPLC at ~1 mL/min; most pharmacopoeial methods. 150 mm is the general default, 250 mm for difficult impurity separations.
- 3.0 mm ID — saves solvent, still works on standard systems.
- 2.1 mm ID — UHPLC and LC-MS (0.2–0.6 mL/min); requires a low-dispersion system.
- Shorter columns (50–100 mm) give faster runs; longer columns give more plates.
4. Choose particle size (and particle type)
- 5 µm — robust, low back-pressure, standard HPLC (≤400 bar).
- 3–3.5 µm — more efficiency at moderate pressure.
- Sub-2 µm fully porous (e.g. Waters ACQUITY UPLC BEH C18 Column, 2.1 x 100 mm, 1.7 µm (
186002352)) — highest efficiency; needs UHPLC pressure capability. - 2.6–2.7 µm core-shell (Phenomenex Kinetex C18 Core-Shell Column, 2.1 x 100 mm, 2.6 µm, 100 Å (
00D-4462-AN), Agilent InfinityLab Poroshell 120 EC-C18 Column, 2.1 x 100 mm, 2.7 µm (695775-902)) — near-sub-2 µm efficiency at lower pressure; a popular way to speed up methods on 600 bar systems.
5. Check pore size, pH and temperature limits
Use 80–130 Å pores for small molecules and ≥300 Å for proteins. Stay inside the column's pH range — silica dissolves at high pH and bonded phases hydrolyse at very low pH — and observe the maximum temperature and pressure on the product page.
Protect the investment
A guard cartridge (guard columns; see how to choose a guard column), an inline filter (fittings & inline filters) and filtered samples (syringe filter selection) are the cheapest way to extend column life.
Matching an existing method
If a method names a specific column, order the same manufacturer part number — search it in the catalog. USP codes help find alternatives, but columns with the same L-code are not automatically interchangeable; see the C18 comparison table and the column cross-reference.
Frequently asked questions
What is the most common HPLC column?
A reversed-phase C18 (USP L1) column is the usual starting point, typically 4.6 x 150 mm or 4.6 x 250 mm with 5 µm particles for conventional HPLC, or 2.1 mm ID with sub-2 µm or core-shell particles for UHPLC and LC-MS.
What is the difference between a C18 and a C8 column?
A C8 phase has shorter alkyl chains than C18, so it is generally less retentive for hydrophobic compounds. Choose C8 when analytes are retained too strongly on C18 or run times are too long.
Does a smaller particle size always give better results?
Smaller particles give higher efficiency but much higher back-pressure. Sub-2 µm columns need a UHPLC system; 2.6–2.7 µm core-shell columns are a common compromise on 600 bar systems, and 5 µm remains the robust choice for standard HPLC.